文件名称:
E H(恩德斯豪斯)Ceragel CPS7272D手册.pdf.pdf
开发工具:
文件大小: 602kb
下载次数: 0
上传时间: 2019-10-31
详细说明:E H(恩德斯豪斯)Ceragel CPS7272D手册.pdfpdf,E H(恩德斯豪斯)Ceragel CPS7272D手册.pdfCeragel CPS72/CPS72D
Data storage of CPS72D
Digital sensors are able to store the following system data in the sensor
Manufacturing data
Serial number
Order code
Date of manufacture
■ Calibration cata
Calibration date
Calibrated offset(operating mode " mv")
%o Slope (operating mode%)
er of calibrations
Serial number of the transmitter used for the last calibration
■ Application data
Temperature application range
Redox application range
Date of first commissioning
Operating hours
These system data can be displayed with the mycom s or the liquiline m cm42 transmitter
Measuring system
A complete measuring system consists of
CPS72 or CPS72D ORP electrode
Transmitter, e.g. Liquisys M CPM223/253 (with Memosens technology for CPS72D
Special measuring cable, e.g. CPK9 or CYK10 Memasens data cable for CPS72D
a Immersion, flow or retractable assembly, e. g. Cleanfit H CPA475
⊙
母
1 CPS72/ CpS72D ORP electrode
2 Gleanfit H CPA475
bly
3 PK9 speciai mea suring cable for electrodes with ToP58 plug-in head)/cyK10 for digital sensors
4 Liquisys l CP253 transmitter
Endress+hau
Ceragel cPs72/CPs72D
Input
Measured variables
Redox potential
Measuring ran
1500.1500mV
Please note the process operating conditions
Installation
Installation instructions
Do not install the electrode upside down. The inclination angle must be at least 15 from the horizontal. A
smaller inclination angle is not permitted as such an inclination results in air cushion forming This might impair
the contact of reference and metal lead
a Make sure that the assemblys threaded connection for the electrode is clean and well running before
installing the electrode
a Hand tighten the electrode( 3 nm)! (Given value only applies to installation in EndreSs+ Hauser assemblies
a Make sure to follow the installation instructions in the operating instructions of the used assembly
15
15
Electrode installation; inclination angie min. 15 from the horizontal
Permitted inclination angle
B Non-permitted inclination angle
Environment
Ambient temperature
Caution
Danger of frost damage
Do not use the electrode at temperatures belOw C/5F
Storage temperature
0..50°C/32..122°F
Ingress protection
TP67:
GSA plug-in head ( witn closed connection syste
TP68:
TOP68 plug-in head(1 m /3.28 ft water column, 50C/122F,168 hi
IP 68
Memosens plug-in head (10 m/32.8 it water column, 25C/77F, 45 days, I M KCI
Endress+hauser
Ceragel CPS72/CPS72D
Process
Process temperature
-15..135°C/5..275°F
Process pressure
0….10bar/0..145psi
Application
Reducing media, e.g. chromate reduction, chlorine dosing in swimming pools
Caution!
Danger of damage to the electrode
Never use the electrode in applications outside the given specifications
Mechanical construction
Design, dimensions
m/ inch
mm/inch
8
3
N=寸o
1210.47
12/047
CPS72 with GSA plug in head
CPS72 with ESa plug in head
I GSA plug-in head, Pg 13.5
7 ESA Plug-in head, P8 13.5
2 Viton O-ring with thrust collar
2 Viton O-ring with thrust collar
Ag/AgCl meta! lead
3Ag/≠ gCl metai lead
4
Bridge electrolyte
4 Bridge eiectrolyte
5
Diaphragm
5 Diaphragm
o Platinum ring
0 Platinum ring
Endress+hauser
Ceragel CPS72 /CPS72D
Design, dimensions CPS72D
6
12/047
CPS72) with Memosens plug-in head
I Memosens plug-in
2 Viton O-ring with thrust collal
3 Ag/AgCl metal
Diaphragm
m ring
Weight
0.1kg/0.2b
Material
Electrode shaft
process gla
Redox measuring element platinum ring
Reference lead
Ag/Agcl
Diaph
ceramic, sterilisable
Process connection
Pg 13.5
Plug-in heads
CPS72
ESA: threaded plug-in head Pg 13.5, TOP 68, 16 bar /232 psi, Ex
threaded plug-in head Pg 13.5
CPS72D: Memosens plug-in head for digital, contactless data transmission, 16 bar/232 psi, Ex or non-EX
Electrolytes
Advanced Gel 3 M KCl AgCl free
Bridge electrolyte
Endress+h
Ceragel CPS72/CPS72D
Certificates and approvals
Ex approval CPS72 (ESA)
aTEXI IG EEX ia g t3/T4/t6
a FM Class I Div. 2, in combination with the liquiline m cM42 and Mycom S CPMI53 transmitters
Ex approval cPs72D
ateX II 1G EEX ia Ic t3/t4/to
FM/ CSA Class I Div. 2, in combination with the Liquiline m cM42 and Mycom S CPM153 transmitters
Note
Ex versions of digital sensors with Memosens technology are indicated by an orange-red ring in the plug-in
head
Biocompatibility
Biocompatibility validated according to
■IS109935:1993
a USP. current revision
TUV certificate toP68 and
Pressure resistance 16 bar, min. triple overpressure safety
ns plug-in head
Electromagnetic compatibility Interference emission and interference immunity complies with En 61326: 1997/Al: 1998
CPSZ2D
Ordering information
Product structure cPs72
Electrode type
Standard version
asuring e
Platinum ring
ft length
120mm/4.72
225mm/8.86
5300mm/14.1
Plug-in head
ESA Plug-in head Pg 13.5, TOP 68, 1o bar/232 psi, Ex
GSA Plug-in head Pg 13.5, DIN coax, non-Ex
CPS72-
complete order cod
Product structure cPS2D
Version
Measuring element
PB Platinurn ring
Shaft length
Pproval
Non-hazardous areas
ATEXIL IG EEx ia Ic t3/4/Tc
CPS72D
complete order code
al pend
Endress+hauser
Ceragel CPS72 /CPS72D
Accessories
Nore
In the following sections, you find the accessories available at the time of issue of this documentation
For information on accessories that are not listed here, please contact your responsible service
Assemblies
■ Cleanfit w cpa150
ually operated, retractable assembly for installation of 120 mm /4.72 pH/ORP
p
Ordering acc. to product structure, see Technical Information(TI183C/07/en)
Cleanfit p cPa471
Compact retractable stainless steel assembly for installation in tanks and pipes, manual or pneumatic
Ordering acc to product structure, see Technical Information(T1217C/07/en)
Cleanfit p cpa472
Compact retractable plastic assembly for installation in tanks and pipes, manual or pneumatic operation,
Ordering acc. to product structure see Technical Information (Ti223C/07/en)
■ Cleanfit p cpa473
Retractable stainless steel process assembly, with ball valve for a particularly safe and reliable separation of
the medium from the environment
Ordering acc. to product structure, see Technical Information(T1344C/07/en
Cleanfit P cPa474
Retractable plastic process assembly, with ball valve for a particularly safe and reliable separation of th
medium from the environment
Ordering acc. to product structure, see Technical Information(T1345C/07/en
0003137
Cleanfit w CPA450
Cleanfit p cpa4zi or 472
Cleanft p cpa473 or 474
Endress+h
Ceragel CPS72/CPS72D
■ Cleanfith cpa475
Retractable assembly for installation in tanks and pipes under sterile conditions
Ordering acc. to product structure, see Technical Information (T1240C/07/eni
■ Unifith cPa442
Process assembly for the food industry, biotechnology and pharmaceutical industry, with EhEdg and 3A
rtificates
Ordering acc to product structure, see Technical Information [TI306C/07/en
Dipfir W CPA
Plastic immersion and installation assembly for open and closed tanks
Ordering acc. to product structure, see Technical Information (TIl 12C/07/en
Cleanfif H CPA475
Unit H CPA442
Dipfit W CPA1!7
Dipfit P CPal
Immersion assembly for pH/ORP electrodes for demanding processes
Ordering acc to product structure, see Technical Information [TI178C/07/en
■ flowfit P cpa240
Flow assembly for pH/ ORP electrodes, for demanding pro
Ordering acc. to product structure, see Technical Information (/07/en
Flowfit W cPa250
Flow assembly for ph/oRP electrod
Ordering acc. to product structure, see Technical Information (T1041C/07/en
Dipfit P CPA140
Flowflt P CPA240
Flowit W CPA250
Endress+hau
Ceragel cPs72/CPs72D
■ Probfith cpa465
Retractable assembly for installation in tanks and pipes under sterile conditions
Ordering acc to product structure, see Technical Intormation (TI146(/07/en
Ecofit CPa640
Process connection adapter and cable set for 120 mm pH/ORP electrodes,
Ordering acC. to product structure, see Technical Information (Ti264C/07/en
Probit H CPA405
Ecofit CPA040
Buffer solutions
Technical buffer solutions for orp electrodes
3+220 mV, pH7.0, 100 ml (0.026 US gal. order no. CPY3-0
+468 mV, pH O 1, 100 ml (0.026 US gal ) order no. CPY3-1
Transmitters
a Liquisys M CPM223/2.53
Transmitter for pH and redox, field or panel-mounted housing,
Hart or ProfiBuS available
Ordering acc to product structure, see Technical Information(TI194C/07/en)
Aycom S CPMI5
Transmitter for pH and redox, one or two channel version, Ex or Non-EX,
Hart or profibus available
Ordering acc to product structure, see Technical Information(t1233C/07/en
■ Liquiline M cm42
Modular two-wire transmitter for ex and non-Ex areas
Harte profibus or foundation fieldbus available
Ordering acc to product structure, see Technical Information(TI381C/07/en
Endress+hauser
(系统自动生成,下载前可以参看下载内容)
下载文件列表
相关说明
- 本站资源为会员上传分享交流与学习,如有侵犯您的权益,请联系我们删除.
- 本站是交换下载平台,提供交流渠道,下载内容来自于网络,除下载问题外,其它问题请自行百度。
- 本站已设置防盗链,请勿用迅雷、QQ旋风等多线程下载软件下载资源,下载后用WinRAR最新版进行解压.
- 如果您发现内容无法下载,请稍后再次尝试;或者到消费记录里找到下载记录反馈给我们.
- 下载后发现下载的内容跟说明不相乎,请到消费记录里找到下载记录反馈给我们,经确认后退回积分.
- 如下载前有疑问,可以通过点击"提供者"的名字,查看对方的联系方式,联系对方咨询.